1. The pointer does not return to zero. (1) If the pointer is bent or loose, it can be corrected with tweezers and tightened after verification. (2) The balance of the hairspring is insufficient, the meshing of the sun gear and the sector gear can be disengaged, and the sun gear shaft is rotated counterclockwise to increase the return torque of the balance spring.


  2. The pointer of the pressure gauge has a beating or sluggishness. (1) The pointer is rubbed against the surface glass or dial. At this point, the pointer can be corrected, the washer under the glass can be thickened or the pointer shaft hole can be twisted larger. (2) The center gear shaft is bent, the shaft diameter is not concentric, and the gear can be removed with a wooden hammer to correct the straightening or straightening with a flat pliers. (3) There is dirt in the meshing of the two gears, and the two gears can be removed for cleaning.


3. A verification point is out of tolerance. According to the requirements of the verification procedure, the indication of the pressure gauge shall be carried out according to the index line marked with numbers. So on which scale the overshoot is found, it is on which scale. Check the matching condition of each part on the scale point, whether the transmission shaft hole is blocked or not; whether the connecting rod is flexible; whether the tooth meshing point is damaged or not, and foreign matter is excluded. If a positive error occurs at a certain point, it often has dirt and burrs due to the tooth engagement point; when a negative error occurs, it is mostly repaired or replaced due to the shape loss or injury of the tooth.


4. The indication value is not stable and the pressure is significantly reduced. At this time, first check whether the connection between the pressure gauge and the pressure gauge checker is leaking. If there is no oil leakage, the calibrator is in good condition, indicating that the spring tube is leaking inside. The scale plate should be removed for further inspection. (1) The base itself has blisters, which gradually leak after long-term use. It is necessary to replace the frame of the same specification, and the low pressure can be repaired. (2) Long-term exposure to pulse pressure, spring tube fatigue or leakage at the seals at both ends. The seals at both ends can be repaired or re-tightened, and the pressure gauge is directly scrapped.


5. The pressure indicates that the value error is not uniform. (1) The deformation of the spring tube is invalid, and the displacement is not proportional to the pressure. The spring tube needs to be replaced. (2) The free end of the spring tube and the gear ratio of the sector gear and the wheel rod are not properly adjusted, and need to be re-inspected and adjusted. (3) The gear clamp plate and the bottom plate are not in the correct position. The splint should be loosened and the splint should be driven in a counterclockwise direction.


6. The gauge pointer does not rotate. When the applied pressure reaches the upper limit of the pressure gauge measurement, the pointer of the pressure gauge remains still or rarely moves, indicating that the pressure is not transmitted to the spring tube. At this time, first check whether the gasket connected to the pressure gauge calibrator blocks the passage at the joint. If there is no problem at the connection, it proves that the pressure gauge is clogged inside, it should be disassembled for cleaning, and then checked.