Before installing the diaphragm pressure gauge, carefully check the model, diaphragm material and measuring range to check whether the diaphragm is bumped. When there is no pressure, the pressure gauge indicates whether it is at zero position, find the cause, re-adjust, and Only after passing the test can it be used. The procedure for calibrating the diaphragm pressure gauge is the same as verifying the conventional pressure gauge, and will not be repeated again. The main points are as follows:
(1) Before the calibration, firstly prepare the matching joint according to the model of the diaphragm device of the diaphragm pressure gauge to be used in the general pressure calibration table. The diaphragm device of the diaphragm pressure gauge is generally flanged. According to the size and type of the flange, the matching pressure fittings can be prepared, and the corresponding flanges and universal pressure gauge joints can be found in the field for processing, which is the most direct and convenient method.
(2) Pay attention to the process of fixing the diaphragm pressure gauge in the standard calibration table: handle it gently, do not touch the diaphragm. When using the wrench to rotate the diaphragm pressure gauge, the wrench must be stuck. The connecting nut of the diaphragm device is not stuck on the universal pressure gauge. Because the universal pressure gauge and the diaphragm device on the instrument are a sealing system, the pressure gauge should not be dismantled or replaced, so as to prevent the working fluid filled inside the system from leaking. The impact on performance and even scrap, especially for new students who have just joined the work, must be explained before the diaphragm pressure gauge is verified.
(3) Result processing: In the process of boosting, if compared with the standard table, if it is out of tolerance, first check whether there is oil leakage, sealing, etc. If there is oil leakage and the seal is not good, it can only be returned to the factory for treatment or scrapped; if not In the above situation, the indicator needle can be removed and the indicator needle can be re-set. The long-term experience is that it is generally reasonable to re-fix the needle at 1/2 of the range when the needle is fixed; if it is still too poor, it can only be degraded or scrapped.